EasyManuals Logo

Keithley 2015 Service Manual

Keithley 2015
158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #129 background imageLoading...
Page #129 background image
Calibration Reference B-3
Command summary
Table B-1 summarizes Model 2015 calibration commands.
Table B-1
Remote calibration command summary
Command Description
:CALibration Calibration root command.
:PROTected All commands in this subsystem are protected by the calibration lock
(except queries and :CODE).
:CODE <up to 8 char. string> Calibration code or password (default: KI002015).
:COUNt? Request the number of times the unit has been calibrated.
:INITiate Initiate calibration.
:LOCK Lock out calibration (opposite of enabling cal with :CODE
command).
:LOCK? Request comprehensive cal lock state. (0 = locked; 1 = unlocked).
:SAVE Save cal constants to EEROM.
:DATE <year>, <month>, <day> Send cal date to 2015.
:DATE? Request cal date from 2015.
:NDUE <year>, <month>, <day> Send next due cal date to 2015.
:NDUE? Request next due cal date from 2015.
:DC DC cal steps.
:STEP0 Rear terminal short step
1
.
:STEP1 Front terminal short circuit.
:STEP2 Open circuit.
:STEP3 <NRf> 10V DC step.
:STEP4 <NRf> –10V DC step.
:STEP5 <NRf> 100V DC step.
:STEP6 <NRf> 1k
4-wire step.
:STEP7 <NRf> 10k
4-wire step.
:STEP8 <NRf> 100k
4-wire step.
:STEP9 <NRf> 1M
4-wire step.
:STEP10 <NRf> 10mA DC step.
:STEP11 <NRf> 100mA DC step.
:STEP12 <NRf> 1A DC step.
NOTES:
1. DC:STEP0, AC:STEP14, and AC:STEP15 are one-time factory calibration points and are valid only in man-
ufacturing calibration mode.
2. Upper-case letters indicated short form of each command. For example, instead of sending
“:CALibration:PROTected:INITiate,” you can send “:CAL:PROT:INIT.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2015 and is the answer not in the manual?

Keithley 2015 Specifications

General IconGeneral
ModelKeithley 2015
CategoryMultimeter
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 300 kHz
Resistance Range100 MΩ
DC Current Range3 A
DC Current Accuracy0.05%
AC Current Range3 A
Display5.5 digit
ConnectivityGPIB
Power Supply100 V to 240 V AC, 50 Hz to 60 Hz

Related product manuals