B-14 Calibration Reference
AC calibration commands
The :AC commands perform comprehensive (user) calibration of the ACV and ACI functions.
Table B-3 summarizes these calibration commands.
Table B-3
AC calibration commands
Command Description
:CALibration
:PROTected
:AC
:STEP1 10mV AC at 1kHz calibration step.
:STEP2 100mV AC at 1kHz calibration step.
:STEP3 100mV AC at 50kHz calibration step.
:STEP4 1V AC at 1kHz calibration step.
:STEP5 1V AC at 50kHz calibration step.
:STEP6 10V AC at 1kHz calibration step.
:STEP7 10V AC at 50kHz calibration step.
:STEP8 100V AC at 1kHz calibration step.
:STEP9 100V AC at 50kHz calibration step.
:STEP10 700V AC at 1kHz calibration step.
:STEP11 100mA AC at 1kHz calibration step.
:STEP12 1A AC at 1kHz calibration step.
:STEP13 2A AC at 1kHz calibration step.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2015 and is the answer not in the manual?

Keithley 2015 Specifications

General IconGeneral
ModelKeithley 2015
CategoryMultimeter
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 300 kHz
Resistance Range100 MΩ
DC Current Range3 A
DC Current Accuracy0.05%
AC Current Range3 A
Display5.5 digit
ConnectivityGPIB
Power Supply100 V to 240 V AC, 50 Hz to 60 Hz