EasyManua.ls Logo

Keithley 2750 - Examples

Keithley 2750
452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-52 Basic DMM Operation Model 2750 Multimeter/Switch System User’s Manual
Examples
One-shot reading, DC volts, no trigger, fastest rate
*RST
:INITiate:CONTinuous OFF;:ABORt
:SENSe:FUNCtion ‘VOLTage:DC’
:SENSe:VOLTage:DC:RANGe 10 // Use fixed range for fastest readings.
:SENSe:VOLTage:DC:NPLC 0.01 // Use lowest NPLC setting for fastest
// readings.
:DISPlay:ENABle OFF // Turn off display to increase speed.
:SYSTem:AZERo:STATe OFF // Disable autozero to increase speed, but may cause
// drift over time.
:SENSe:VOLTage:DC:AVERage:STATe OFF // Turn off filter for speed.
:TRIGger:COUNt 1
:READ?
(Enter reading)
One-shot reading, DC volts, bus trigger, auto ranging
*RST
:INITiate:CONTinuous OFF;:ABORt
:TRIGger:SOURce BUS
:SENSe:FUNCtion ‘VOLTage:DC’
:SENSe:VOLTage:DC:RANGe:AUTO ON
:TRIGger:COUNt 1
:INITiate
*TRG -or- GET // Triggers reading (GET is a GPIB general bus command).
:SENSe:DATA:FRESh?
(Enter reading)
One-shot reading, external trigger, auto delay enabled
*RST
:INITiate:CONTinuous OFF;:ABORt
:TRIGger:SOURce EXTernal
:TRIGger:DELay:AUTO ON // Note: Auto trigger delay only takes effect with
// trigger source set for BUS or EXTernal.
:SENSe:FUNCtion ‘VOLTage:DC’
:SENSe:VOLTage:DC:RANGe:AUTO ON
:INITiate
(external trigger)
:SENSe:DATA:FRESh?
(enter reading) // This step will time out if the trigger hasn’t occurred.
2750-900-01.book Page 52 Wednesday, August 3, 2011 7:56 AM

Table of Contents

Other manuals for Keithley 2750

Related product manuals