Model 2750 Multimeter/Switch System User’s Manual SCPI Signal Oriented Commands 13-5
FETCh?
Description This command requests the latest post-processed reading. After sending
this command and addressing the Model 2750 to talk, the reading is sent
to the computer. This command does not affect the instrument setup.
This command does not trigger a measurement. The command simply
requests the last available reading.
FETCh? can also be used to return more than one reading. When return-
ing more than one reading, the readings are automatically stored in the
buffer.
In order to return multiple reading strings, continuous initiation must be
disabled (INIT:CONT OFF) so that the sample count
(SAMPle:COUNt), which specifies the number of measurements to be
performed, can be set >1. After INITiate is sent to trigger the measure-
ments, FETCh? will return the reading strings.
FETCh? is automatically asserted when the READ? or MEASure? com-
mand is sent.
NOTES FETCh? can repeatedly return the same reading. Until there is a new reading,
this command continues to return the old reading.
When an instrument setting that is relevant to the readings in the sample buffer
is changed, the FETCh? command will cause error -230 (data corrupt or stale)
or a bus time-out to occur. To get FETCh? working again, a new reading must
be triggered.
READ?
Description This command is typically used with the instrument in the “one-shot”
measurement mode to trigger and acquire a specified number of read-
ings. The SAMPle:COUNt command is used to specify the number of
readings (see Triggering commands on page 8-16). Note that with sam-
ple count >1, the readings are stored in the buffer.
When this command is sent, the following commands execute in the
order they are presented:
ABORt
INITiate
FETCh?
When ABORt is executed, the instrument goes into the idle state if con-
tinuous initiation is disabled. If continuous initiation is enabled, the
operation re-starts at the beginning of the Trigger Model.
2750-900-01.book Page 5 Wednesday, August 3, 2011 7:56 AM

Table of Contents

Other manuals for Keithley 2750

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2750 and is the answer not in the manual?

Keithley 2750 Specifications

General IconGeneral
InterfaceGPIB, RS-232, Ethernet
Switching SpeedUp to 500 channels/second
Maximum Voltage300V
Operating Temperature0°C to 50°C
Measurement FunctionsDCV, ACV, DCI, ACI, 2-wire Resistance, 4-wire Resistance, Frequency, Temperature
Resolution6.5 digits
Current1A
Maximum Current1A