Model 2750 Multimeter/Switch System User’s Manual Status Structure 11-5
Programming and reading registers
Programming enable registers
The only registers that can be programmed by the user are the enable registers. All other
registers in the status structure are read-only registers. The following explains how to
ascertain the parameter values for the various commands used to program enable registers.
The actual commands are covered later in this section (Table 11-2 and Table 11-5).
A command to program an event enable register is sent with a decimal parameter value
that determines the desired state (0 or 1) of each bit in the appropriate register. The bit
positions of the register (Figure 11-2) indicate the binary parameter value. For example, if
you wish to set bits B4, B3, and B1, the binary value would be 11010 (where B4=1, B3=1,
B1=1, and all other bits are 0).
The binary value is then converted to its decimal equivalent:
Binary 11010 = Decimal 26
Figure 11-2 includes the decimal weight for each register bit. To set bits B4, B3, and B1,
the decimal parameter value would be the sum of the decimal weights for those bits
(16 + 8 + 2 = 26).
Figure 11-2
16-bit status register
(2
7
)(2
6
)(2
5
)(2
4
)(2
3
)
(2
2
)
(2
1
)(2
0
)
(2
15
)(2
14
)
(2
13
)(2
12
)(2
11
)(2
10
)(2
9
)(2
8
)
32768
16384
Bit Position
Binary Value
Decimal
Weights
B7
B6
B5
B4
B3
B2
B1
B0
0/1
0/1 0/1
0/1 0/1 0/10/10/1
128
64
32
16
8
4
2
1
B15 B14 B13 B12 B11 B10 B9 B8
8192 4096 2048 1024 512 256
0/1
0/1 0/1
0/1 0/1 0/10/10/1
Bit Position
Binary Value
Decimal
Weights
B. Bits 8 through 15
A. Bits 0 through 7
2750-900-01.book Page 5 Wednesday, August 3, 2011 7:56 AM

Table of Contents

Other manuals for Keithley 2750

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2750 and is the answer not in the manual?

Keithley 2750 Specifications

General IconGeneral
InterfaceGPIB, RS-232, Ethernet
Switching SpeedUp to 500 channels/second
Maximum Voltage300V
Operating Temperature0°C to 50°C
Measurement FunctionsDCV, ACV, DCI, ACI, 2-wire Resistance, 4-wire Resistance, Frequency, Temperature
Resolution6.5 digits
Current1A
Maximum Current1A

Related product manuals