Model 2750 Multimeter/Switch System User’s Manual Scanning 7-19
Method 1:
1. Use the CLOSE key or the and keys to close the channel that you want to be
the monitor.
2. Press SHIFT and then MONITOR (MON annunciator turns on).
Method 2:
1. If a channel is closed, press OPEN to open it.
2. Press SHIFT and then MONITOR.
3. Use the , ,
Δ , and keys to display the monitor channel (i.e., MONITOR
101), and press ENTER. The monitor channel closes and the MON annunciator
turns on.
To disable monitor, again press SHIFT and then MONITOR (MON annunciator turns off).
Once enabled, you can change the monitor channel using the CLOSE key or the and
keys. If you open the monitor channel, the monitor does not disable but it does become
inactive (MON annunciator turns off). When a channel is closed, monitor becomes active
(MON annunciator turns on).
While in the normal measurement state, the present monitor channel dictates which chan-
nel in the scan list is the monitor. Therefore, if you change the monitor channel, the scan
list monitor channel also changes.
When you change the monitor channel while in the normal measurement state, the instru-
ment setup does not change. If you want the monitor channel to assume the setup of the
scan list channel, you must disable the monitor and then re-enable it.
Auto channel configuration
Auto channel configuration allows you to recall scan list setups. With auto channel config-
uration enabled, a closed channel assumes the scan list setup. With this feature, you can
inspect the channel setups of the scan, or manually scan channels. When a scan channel is
disabled (not in scan list), it cannot be closed with auto channel configuration enabled.
As with normal operation, when you use the , , or CLOSE keys to close a channel (or
channel pair), any other closed channels are first opened.
Perform the following steps to enable or disable auto channel configuration.
1. Press SHIFT and then SETUP.
2. Use the
Δ or key to display the auto configuration (CH AUTOCFG) setting; N
(no) or Y (yes).
3. Press the key to place the cursor on the present setting (N or Y), and press the
Δ
or
key to change the setting.
4. Press ENTER to return to the normal measurement state.
2750-900-01.book Page 19 Wednesday, August 3, 2011 7:56 AM

Table of Contents

Other manuals for Keithley 2750

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2750 and is the answer not in the manual?

Keithley 2750 Specifications

General IconGeneral
InterfaceGPIB, RS-232, Ethernet
Switching SpeedUp to 500 channels/second
Maximum Voltage300V
Operating Temperature0°C to 50°C
Measurement FunctionsDCV, ACV, DCI, ACI, 2-wire Resistance, 4-wire Resistance, Frequency, Temperature
Resolution6.5 digits
Current1A
Maximum Current1A