Model 2750 Multimeter/Switch System User’s Manual Triggering 8-11
C. For the first pass through the model, the scanner does not wait at point B for a trig-
ger. Instead, it closes the first channel.
D. After the relay settles, the Model 7002 outputs a Channel Ready pulse. Since the
instrument is programmed to scan 400 channels, operation loops back up to point
B, where it waits for an input trigger.
E & F. Model 2750 operation is at point A waiting for a trigger. The output Channel
Ready pulse from the Model 7002 triggers the Model 2750 to measure DUT #1
(point E). After the measurement is complete, the Model 2750 outputs a comple-
tion pulse (point F) and then loops back to point A where it waits for another input
trigger.
The trigger applied to the Model 7002 from the Model 2750 closes the next channel in the
scan. This triggers the Model 2750 to measure the next DUT. The process continues until
all 400 channels are scanned, measured, and stored in the buffer.
External triggering with BNC connections
An adapter cable is available to connect the micro-DIN Trigger Link of the Model 2750 to
instruments with BNC trigger connections. The Model 8503 DIN to BNC Trigger Cable
has a micro-DIN connector at one end and two BNC connectors at the other end. The BNC
cables are labeled VMC (trigger line 1) and EXT TRIG (trigger line 2).
Figure 8-9 shows how a Keithley Model 220 Current Source can be connected to the Trig-
ger Link of the Model 2750 using the adapter cable. When used with the STEP mode of
the Model 220, you can perform synchronized source-measure operations without the use
of a computer. Whenever the Model 220 receives a trigger from the Model 2750, it will
step to the next current source value.
2750-900-01.book Page 11 Wednesday, August 3, 2011 7:56 AM

Table of Contents

Other manuals for Keithley 2750

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2750 and is the answer not in the manual?

Keithley 2750 Specifications

General IconGeneral
InterfaceGPIB, RS-232, Ethernet
Switching SpeedUp to 500 channels/second
Maximum Voltage300V
Operating Temperature0°C to 50°C
Measurement FunctionsDCV, ACV, DCI, ACI, 2-wire Resistance, 4-wire Resistance, Frequency, Temperature
Resolution6.5 digits
Current1A
Maximum Current1A

Related product manuals