5-10 Rel, Math, Ratio, Channel Average, dB Model 2750 Multimeter/Switch System User’s Manual
mX+b rel
The mX+b function can be used to manually establish a rel value. To do this, set the scale
factor (M) to 1 and set the offset (b) to the rel value. Each subsequent reading will be the
difference between the actual input and the rel value (offset).
Percent
This math function determines percent deviation from a specified reference value. The
percent calculation is performed as follows:
where: Input is the normal display reading.
Reference is the user entered constant.
Percent is the displayed result.
NOTE When using Rel, the rel’ed reading of the input signal is used by the percent
calculation.
Percent configuration
1. Press SHIFT and then MATH to display the math menu.
2. Press the RANGE
Δ or key to display “PERCENT,” and press ENTER to dis-
play the present reference value:
REF +1.000000
Δ (factory default)
3. Key in the reference value. The and keys control cursor position, and the
Δ
and
keys increment/decrement the digit value. To change range, place the cursor
on the multiplier and use the
Δ and keys (m = × 0.001, Δ = ×1, K = × 1000,
and M = × 1,000,000). With the cursor on the polarity sign, the
Δ and keys tog-
gle polarity.
4. Press ENTER. The MATH annunciator will turn on, and the result of the calcula-
tion will be displayed. Note that the calculation will be applied to all measurement
functions.
5. To disable mX+b, again press SHIFT and then MATH. The MATH annunciator
will turn off.
NOTE The result of the percent calculation is positive when the input exceeds the refer-
ence and negative when the input is less than the reference.
The result of the percent calculation may be displayed in exponential notation.
For example, a displayed reading of +2.500E+03% is equivalent to 2500%
(2.5K%).
Percent
Input Reference
Reference
-------------------------------------------
100%×=
2750-900-01.book Page 10 Wednesday, August 3, 2011 7:56 AM

Table of Contents

Other manuals for Keithley 2750

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2750 and is the answer not in the manual?

Keithley 2750 Specifications

General IconGeneral
InterfaceGPIB, RS-232, Ethernet
Switching SpeedUp to 500 channels/second
Maximum Voltage300V
Operating Temperature0°C to 50°C
Measurement FunctionsDCV, ACV, DCI, ACI, 2-wire Resistance, 4-wire Resistance, Frequency, Temperature
Resolution6.5 digits
Current1A
Maximum Current1A

Related product manuals