7-26 Scanning Model 2750 Multimeter/Switch System User’s Manual
Remote programming — scanning
NOTE Scanning examples (remote programming and front panel operation) are pro-
vided at the end of this section.
Trigger model
The trigger model for bus operation is shown in Figure 7-2. Bus operation is similar to
front panel SCAN operation, with the following significant differences:
Idle — The instrument goes into the idle state (measurements halted) after the last scan
channel is measured. For front panel operation, the instrument stays in idle until the next
scan is started. For bus operation, the instrument will not stay in idle unless continuous ini-
tiation is disabled. There are two commands to disable continuous initiation:
INITitate:CONTinuous OFF ' Disable continuous initiation.
*RST ' Restore *RST defaults.
The instrument will remain in idle until it receives an initiate command. Typical com-
mands to initiate one scan cycle include:
INITiate ' Initiate one scan cycle.
READ? ' Initiate one scan cycle and request “sample” readings.
More information on using these commands is provided by Reference c that follows
Table 7-1.
Control sources — For bus operation, there are two additional control sources: Bus and
Manual. For the Bus control source, scan operation is controlled by bus triggers (i.e.,
*TRG) or by using the TRIG key. For the Manual control source, event detection is con-
trolled solely by the TRIG key. Note that the instrument has to be in local in order to use
the TRIG key. The LOCAL key takes the instrument out of remote.
Trigger and sample counters — For front panel SCAN operation, the number of chan-
nels in the scan list and the programmed reading count automatically sets the trigger and
sample counters. For remote operation, these two counters are set by the
TRIGGer:COUNt and SAMPle:COUNt commands.
NOTE To set sample count >1, continuous initiation must be disabled (see “Idle”).
Note that only sample count readings are stored in the buffer. See Section 8 for
detailed information on the trigger model.
2750-900-01.book Page 26 Wednesday, August 3, 2011 7:56 AM

Table of Contents

Other manuals for Keithley 2750

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2750 and is the answer not in the manual?

Keithley 2750 Specifications

General IconGeneral
InterfaceGPIB, RS-232, Ethernet
Switching SpeedUp to 500 channels/second
Maximum Voltage300V
Operating Temperature0°C to 50°C
Measurement FunctionsDCV, ACV, DCI, ACI, 2-wire Resistance, 4-wire Resistance, Frequency, Temperature
Resolution6.5 digits
Current1A
Maximum Current1A

Related product manuals