Model 2750 Multimeter/Switch System User’s Manual Scanning 7-25
a. Press the Δ or ∇ key to enable or disable low limit 1 (LLIM1 SCAN:N/Y),
and press ENTER.
b. Press the
Δ or ∇ key to enable or disable high limit 1 (HLIM1 SCAN:N/Y),
and press ENTER.
c. Press the
Δ or ∇ key to enable or disable low limit 2 (LLIM2 SCAN:N/Y),
and press ENTER.
d. Press the
Δ or ∇ key to enable or disable high limit 2 (HLIM2 SCAN:N/Y),
and press ENTER.
3. Finish configuring the scan by performing Step 4 and Step 5 of the “Advanced scan
setup procedure,” page 7-16.
NOTE For a remote programmed monitor scan, use the ROUTe:MONitor:POINts com-
mand to specify the number of channels to scan (Table 7-1).
4. While in the normal measurement state, select and enable the monitor channel as
explained in “(see Scan configuration) — Monitor channel,” page 7-18.
When the reading limit for the monitor channel is reached, the scan will be trig-
gered to start. When the monitor channel is scanned, the display will show the
reading that triggered the scan.
If the reading limit event is still present on the monitor channel when the scan fin-
ishes, the scan will be triggered to run again. Note that the scan can also be run by
pressing STEP or SCAN.
5. To disable the monitor scan, perform the following steps:
a. To disable monitor, press SHIFT and then MONITOR (MON annunciator
turns off).
b. To disable limits, press SHIFT and then ON/OFF. Press
Δ or ∇ to display
“LIMITS: OFF” and press ENTER.