Model 2750 Multimeter/Switch System User’s Manual Scanning 7-25
a. Press the Δ or key to enable or disable low limit 1 (LLIM1 SCAN:N/Y),
and press ENTER.
b. Press the
Δ or key to enable or disable high limit 1 (HLIM1 SCAN:N/Y),
and press ENTER.
c. Press the
Δ or key to enable or disable low limit 2 (LLIM2 SCAN:N/Y),
and press ENTER.
d. Press the
Δ or key to enable or disable high limit 2 (HLIM2 SCAN:N/Y),
and press ENTER.
3. Finish configuring the scan by performing Step 4 and Step 5 of the Advanced scan
setup procedure,” page 7-16.
NOTE For a remote programmed monitor scan, use the ROUTe:MONitor:POINts com-
mand to specify the number of channels to scan (Table 7-1).
4. While in the normal measurement state, select and enable the monitor channel as
explained in “(see Scan configuration)Monitor channel,” page 7-18.
When the reading limit for the monitor channel is reached, the scan will be trig-
gered to start. When the monitor channel is scanned, the display will show the
reading that triggered the scan.
If the reading limit event is still present on the monitor channel when the scan fin-
ishes, the scan will be triggered to run again. Note that the scan can also be run by
pressing STEP or SCAN.
5. To disable the monitor scan, perform the following steps:
a. To disable monitor, press SHIFT and then MONITOR (MON annunciator
turns off).
b. To disable limits, press SHIFT and then ON/OFF. Press
Δ or to display
“LIMITS: OFF” and press ENTER.
2750-900-01.book Page 25 Wednesday, August 3, 2011 7:56 AM

Table of Contents

Other manuals for Keithley 2750

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2750 and is the answer not in the manual?

Keithley 2750 Specifications

General IconGeneral
InterfaceGPIB, RS-232, Ethernet
Switching SpeedUp to 500 channels/second
Maximum Voltage300V
Operating Temperature0°C to 50°C
Measurement FunctionsDCV, ACV, DCI, ACI, 2-wire Resistance, 4-wire Resistance, Frequency, Temperature
Resolution6.5 digits
Current1A
Maximum Current1A