1-20 Getting Started Model 2750 Multimeter/Switch System User’s Manual
Saving and restoring setups
Saving a user setup
1. Configure Model 2750 for the desired measurement application.
2. Press SHIFT and then SAVE to access the save setup menu.
3. Press to place the cursor on the present setup (SAV0, SAV1, or SAV2).
4. Use the
Δ or key to display the desired setup and press ENTER. The instrument
returns to the normal measurement state.
WARNING If you make firmware upgrades you will lose all your saved settings.
Saving a power-on setup
1. Configure Model 2750 for the desired measurement application.
2. Press SHIFT and then SAVE to access the save setup menu.
3. Press the
Δ key to display the present power-on (PWR-ON) setup: FACT, *RST,
SAV0, SAV1, or SAV2.
4. Press to place the cursor on the present power-on setup.
5. Use the
Δ or key to display the desired setup and press ENTER. The instrument
returns to the normal measurement state.
Restoring a setup
1. Press SHIFT and then SETUP to access the restore setup menu.
2. Press to place the cursor on the present RESTORE setup (FACT, *RST, SAV0,
SAV1, or SAV2).
3. Use the
Δ or key to display the desired setup and press ENTER. The instrument
returns to the normal measurement state.
NOTE If the settings for a user setup or power-on setup do not match the switching
module types presently installed in the Model 2750, error +520 (Saved setup
scancard mismatch) occurs when the setup is recalled. The scan list will reset to
the factory defaults and all channels will open. However, the saved setup is still
retained in memory and can be restored when the matching switching module is
later installed.
2750-900-01.book Page 20 Wednesday, August 3, 2011 7:56 AM

Table of Contents

Other manuals for Keithley 2750

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2750 and is the answer not in the manual?

Keithley 2750 Specifications

General IconGeneral
InterfaceGPIB, RS-232, Ethernet
Switching SpeedUp to 500 channels/second
Maximum Voltage300V
Operating Temperature0°C to 50°C
Measurement FunctionsDCV, ACV, DCI, ACI, 2-wire Resistance, 4-wire Resistance, Frequency, Temperature
Resolution6.5 digits
Current1A
Maximum Current1A