Model 2750 Multimeter/Switch System User’s Manual Triggering 8-5
Device action
The primary device action is a measurement. However, the device action block could
include the following additional actions (Figure 8-2):
Figure 8-2
Device action
Filtering — If the repeating filter is enabled, the instrument samples the specified
number of reading conversions to yield single filtered reading. Only one reading
conversion is performed if the filter is disabled, or after the specified number of
reading conversions for a moving average filter is reached. After a reading (Rdg) is
procured, operation proceeds to Channel Closure.
Channel Closure — When scanning, the last device action is channel control. The
Model 2750 opens the presently closed channel, and then closes the next channel in
the scan.
Output trigger
After the device action, an output trigger occurs and is available at the rear panel Trigger
Link connector. This trigger can be used to trigger another instrument to perform an oper-
ation (e.g., select the next channel for an external scan).
Rdg Chan
Filter
DEVICE ACTION
To Output Trigger
Block of Figure 8-1
From Delay Block
of Figure 8-1
2750-900-01.book Page 5 Wednesday, August 3, 2011 7:56 AM

Table of Contents

Other manuals for Keithley 2750

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2750 and is the answer not in the manual?

Keithley 2750 Specifications

General IconGeneral
InterfaceGPIB, RS-232, Ethernet
Switching SpeedUp to 500 channels/second
Maximum Voltage300V
Operating Temperature0°C to 50°C
Measurement FunctionsDCV, ACV, DCI, ACI, 2-wire Resistance, 4-wire Resistance, Frequency, Temperature
Resolution6.5 digits
Current1A
Maximum Current1A

Related product manuals