7-2 Scanning Model 2750 Multimeter/Switch System User’s Manual
Scanning fundamentals
The Model 2750 can scan the channels of up to five installed Keithley switching modules.
Each scan channel can have its own unique setup. Aspects of operation that can be
uniquely set for each channel include function, range, rate, AC bandwidth, rel, filter, dig-
its, math, Ω offset compensation, TEMP transducers, limits, channel average, channel
ratio, and volts dB.
NOTE Readings for scanned channels are automatically stored in the buffer. With
buffer auto clear enabled (which is the default), the buffer clears when the scan
is started. When disabled, scan readings are appended to the buffer. The
TRACe:CLEar:AUTO <b> command is used to enable/disable buffer auto clear
(Section 6).
A pseudocard can be installed in an empty slot. With the 7700 pseudocard
installed, the instrument will operate as if a Model 7700 switching module is
installed in the slot. This allows the user to configure a scan and exercise its
operation before the switching module is installed in the Model 2750. Use the
following commands to install 7700 pseudocards in empty slots:
SYSTem:PCARd1 C7700 ‘Install’ 7700 pseudocard in slot 1.
SYSTem:PCARd2 C7700 ‘Install’ 7700 pseudocard in slot 2.
SYSTem:PCARd3 C7700 ‘Install’ 7700 pseudocard in slot 3.
SYSTem:PCARd4 C7700 ‘Install’ 7700 pseudocard in slot 4.
SYSTem:PCARd5 C7700 ‘Install’ 7700 pseudocard in slot 5.
Pseudocards for other switching modules can instead be installed. Details on
installing other pseudocards are provided in Section 2.
There are SCPI commands to query the capabilities of the installed switching modules.
For example, the following queries are provided to determine which channels are available
for volts/2-wire ohms, and which channels are available for amps. Note that the returned
values for the Model 7700 switching module are shown in parenthesis.
SYSTem:CARD:VCHannel[:STARt]? <NRf> ‘ Query first volt/Ω2 channel (7700; “1”).
SYSTem:CARD:VCHannel:END? <NRf> ‘ Query last volt/Ω2 channel (7700; “20”).
SYSTem:CARD:ACHannel[:STARt]? <NRf> ‘ Query first amps channel (7700; “21”).
SYSTem:CARD:ACHannel:END? <NRf> Query last amps channel (7700; “22”).
where: <NRf> = slot number (1 to 5)
All the commands to query switching module capabilities are covered in Table 15-7.
NOTE The Model 2700 uses a different syntax to specify a slot for the :CARD com-
mand. Instead of using the <NRF> parameter, the slot number is included with
the :CARD command word (i.e., SYST:CARD1:VCH:END?). In order to make
the Model 2750 compatible with Model 2700 operation, the :CARD command
for the Model 2750 will accept the 2700 syntax for slots 1 and 2.
2750-900-01.book Page 2 Wednesday, August 3, 2011 7:56 AM

Table of Contents

Other manuals for Keithley 2750

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2750 and is the answer not in the manual?

Keithley 2750 Specifications

General IconGeneral
InterfaceGPIB, RS-232, Ethernet
Switching SpeedUp to 500 channels/second
Maximum Voltage300V
Operating Temperature0°C to 50°C
Measurement FunctionsDCV, ACV, DCI, ACI, 2-wire Resistance, 4-wire Resistance, Frequency, Temperature
Resolution6.5 digits
Current1A
Maximum Current1A