7-12 Scanning Model 2750 Multimeter/Switch System User’s Manual
There are two scan configurations: simple and advanced. When you configure the simple
scan, the instrument uses the present instrument setup for each channel in the scan. For the
advanced scan, each channel can have its own unique setup. As explained in “Trigger
models,” page 7-4, there is a user-set delay (auto or manual) that is in effect for both the
simple and advanced scan.
Channel setup considerations
Rel — In order to use an acquired rel value for an advanced scan channel, the rel value has
to be acquired with the instrument in the normal measurement state. Details to set rel for
scan channels are provided in “Relative,” page 5-2. Scanning examples (front panel and
remote programming) at the end of this section demonstrate how to set rel values for scan
channels.
Filter — The moving filter cannot be used in a scan; only the repeat filter can be used. If
you configure a channel (or channels) to use the moving filter, the filter will be off when
the scan is run. See Section 4 for details on filter.
NOTE When in the scan setup menu, use the edit keys (, ,
Δ , and ) to make
selections and set values. Displayed selections and settings are entered by press-
ing the ENTER key.
Saving the configured scan
The configured scan can be saved in a user-saved setup (SAV0, SAV1, or SAV2). For a
front panel configured scan, the timer state (NO or YES), timer value, and reading count
value are also saved. However, if the settings for a user setup or power-on setup do not
match the switching module type presently installed, error +520 (Saved setup scancard
mismatch) occurs when the setup is recalled. The scan resets to the factory default set-
tings, and all channels will open. The saved setup is still retained in memory and can be
restored when the matching switching module is later installed.
There is an anomaly with the displayed front panel CONFIG menu when a saved front
panel scan is recalled. The reading count value in the menu may not reflect the actual read-
ing count of the scan. For example:
Assume a Model 7700 module is configured for a 10-channel scan and a reading count of
30. For this configuration, the instrument will scan through the scan list three times. Now
assume the scan setup is saved in SAV1, and SAV1 is the power-on default. After cycling
power, press SCAN or STEP to run the scan. The scan will run properly. That is, the 10
channels will be scanned three times. However, if you check the reading count in the front
panel CONFIG menu, the reading count will be 10 (instead of 30). This is because the
scan list length “suggests” the reading count value during the setup process.
NOTE Saving and recalling user-setups are covered in Section 1 (see Defaults and user
setups).
2750-900-01.book Page 12 Wednesday, August 3, 2011 7:56 AM

Table of Contents

Other manuals for Keithley 2750

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2750 and is the answer not in the manual?

Keithley 2750 Specifications

General IconGeneral
InterfaceGPIB, RS-232, Ethernet
Switching SpeedUp to 500 channels/second
Maximum Voltage300V
Operating Temperature0°C to 50°C
Measurement FunctionsDCV, ACV, DCI, ACI, 2-wire Resistance, 4-wire Resistance, Frequency, Temperature
Resolution6.5 digits
Current1A
Maximum Current1A

Related product manuals