8-14 Triggering Model 2750 Multimeter/Switch System User’s Manual
Figure 8-10
Trigger model (remote operation)
Control
Source
Event
Detection
No
START
:INIT (:IMM)
or
:INIT:CONT ON
?
Yes
:ABOrt
*RCL 0
:SYST:PRES
*RST
:INIT (:IMM)
or
:INIT:CONT ON
?
No
Yes
Idle
and
Initiate
:Trigger:Source Immediate
:Trigger:Source External
:Trigger:Source Timer
:Trigger:Source Manual
:Trigger:Source BUS
:Trigger:Delay <n>
:Trigger:Delay:AUTO <b>
Another
Trigger
?
Output
Trigger
Another
Sample
?
Device
Action
Yes
No
Yes
No
:Trigger:Signal
:Sample:Count <n>
:Trigger:Count<n> | INFinity
Timer >
Delay
?
Timer
Enabled
?
Yes
No
Timer
Timer
Bypass
Yes
No
Delay
(Auto or Manual)
2750-900-01.book Page 14 Wednesday, August 3, 2011 7:56 AM

Table of Contents

Other manuals for Keithley 2750

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2750 and is the answer not in the manual?

Keithley 2750 Specifications

General IconGeneral
InterfaceGPIB, RS-232, Ethernet
Switching SpeedUp to 500 channels/second
Maximum Voltage300V
Operating Temperature0°C to 50°C
Measurement FunctionsDCV, ACV, DCI, ACI, 2-wire Resistance, 4-wire Resistance, Frequency, Temperature
Resolution6.5 digits
Current1A
Maximum Current1A