115
3.14 Collimator Assessment
Chapter 3
5
Place the (4) pennies on the edge of the light eld as shown in the picture below. Make sure
not to cover the cells used for AEC.
6
Replace the compression paddle and ensure that it is 4.2 cm above the breast support table.
Use the digital readout to ensure 4.2 cm.
7
Place the quarter up against the chest wall side of the compression paddle.
8
Place 2cm of acrylic on top of the paddle. Note the quarter and penny on the chest wall side
are o center not to interfere with the AEC. The acrylic should extend past the chest wall to
ensure complete coverage.
Quarter Coin
9
Expose the cassettes using AEC and the most commonly used kVp.
10
Process the cassettes and print the resulting images. Make sure to print out at true size.