3.1 Testing Summary
3.1.2 Testing Frequency
All tests contained in this section are to be carried out a) initially upon installation prior to patient examination, b) an-
nually thereafter, c) after any major repair, upgrade, or modication on a component of the imaging chain which may
a󰀨ect the performance to be tested, and d) when problems are suspected.
Some or all of the tests in this section may be required as part of a Mammography Equipment Evaluation (MEE) as
directed by MQSA and recommended by ACR.
Examples of major changes or repairs that would call for an MEE include, but are not limited to:
Replacement of such component of a mammography X-ray unit as X-ray tube, collimator, lter, AEC or AEC sen-
sor.
Replacement of such part of the CR system as optical unit, power supplies, system control board, or A/D board.
Contact your service representative for detailed information about the changes or repairs.