3.18 Inter-Plate Consistency
Chapter 3
3.18.7 Documentation
Record the results in the Table MP-15 (SNR) (Inter-Plate Consistency) and the Table MP-S (MEDICAL PHYSICIST'S
MAMMOGRAPHY QC TEST SUMMARY).
HINT
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
The following is an informative guide to an alternative procedure for Inter-Plate Consistency test. Since S value is a direct
measure of the pixel value of an acquired image (before such image post-processing as Automatic Gradation Processing)
using the REGIUS plate, any inter-plate variation in the response characteristics results in variation in S values.
Test Objective
To ensure that each imaging plate in use provides a uniform response to radiation under standardized conditions by
checking their S value consistency.
Frequency
Initially at installation
ANNUALLY
Before introducing new REGIUS plates or cassettes
Required Equipment
1cm of PMMA (acrylic) phantom
All REGIUS cassettes in use
Procedure
kVp: most commonly used clinical setting
mAs: most commonly used clinical setting; or AEC setting
Compress paddle on top of the 1cm of acrylic
Select the appropriate patient and exam tag
Calculate the mean S value of the reviewed images.
IMPORTANT
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
This test is sensitive to variations in beam quality (kVp, target and lter material). Use an AEC mode that varies only
mAs, and does not allow variations in kVp, target and lter material.
This test is also sensitive to variations in the patient support and anti-scatter grid.
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
Performance Criteria and Corrective Action
Performance Criteria: The reported S value of the individual images should be with in +/- 15% from the mean value.
Corrective Action: Ensure all images evaluated were taken in the same mode under the same X-ray conditions. Erase all
cassettes and repeat the test taking care that cassettes not intended for direct exposure are stored in an area not sub-
ject to scatter radiation.
If a particular cassette and plate pair produces a value which falls far outside the performance criteria, remove that pair
and repeat the test. If the new results pass the test, patient examination may continue without using that cassette and
plate pair.
The source of the problem shall be identied and corrective action shall be taken within 30 days of the test date. Patient
examination may continue during this period.
Documentation
Record the results in the Table MP-15 (Inter-Plate Consistency (S value)) and the Table MP-S (MEDICAL PHYSICIST'S
MAMMOGRAPHY QC TEST SUMMARY).
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

Table of Contents

Other manuals for Konica Minolta CS-3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta CS-3 and is the answer not in the manual?

Konica Minolta CS-3 Specifications

General IconGeneral
BrandKonica Minolta
ModelCS-3
CategoryMedical Equipment
LanguageEnglish

Related product manuals