3.13 AEC Performance Checks
Chapter 3
3.13.2 Thickness Tracking
z
Test Objective
To evaluate the thickness tracking performance of the Automatic Exposure Control (AEC) system.
z
Frequency
Initially before patients are imaged
ANNUALLY
As part of an MEE
z
Required Equipment
2, 4, and 6 cm of PMMA (acrylic) phantom
Al sheet (Thickness: 0.2 mm, Size: 60 mm × 60 mm or larger)
Imaging cassette
z
Procedure (for CS-3)
1
Place PMMA (acrylic) phantom of thickness 2 cm on the breast support table.
2
Position the compression paddle as close as possible to the phantom surface.
3
Move the AEC sensor to its most forward position (closest to the chest side).