3.12 S value Response Indicator
3.12.1 Test Objective
To ensure the displayed S value of the CR Image is within specication.
3.12.2 Frequency
Initially before patients are imaged
ANNUALLY
After major repair, upgrade, or modication of the REGIUS Model 190/210/110HQ or the mammography X-ray unit,
and whenever its calibration problem is suspected
3.12.3 Required Equipment
Calibrated dose meter
REGIUS cassette
3.12.4 Procedure
1
Position the detector of the dose meter slightly o󰀨-center and approximately 4 cm from the
chest wall edge of the breast support table. Position the compression paddle as close as pos-
sible to the detector. Place a REGIUS cassette on the breast support table.
2
Select the Mammo S value exam tag.
HINT
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
Use the same cassette for all exposures.
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
3
Expose with 28kVp in the standard mammography mode (Mo/Mo) and the lowest mAs setting
possible.