3.11 Ghost Image Evaluation
4
Remove the RMI-156 phantom and replace it with the 2cm of PMMA (acrylic) phantom.
5
Register the cassette under the Erase 2 exam tag.
6
Expose the cassette using 28kVp 40mAs.
7
Process the cassette and view the acquired image.
3.11.5 Performance Criteria and Corrective Action
Performance Criteria: The Erase 2 image should be free of any visible indication of the RMI phantom from the previ-
ous exposure.
Corrective Action: If a latent image is detected the REGIUS system will need to have service to address the issue.
The source of the problem shall be identied and corrective action shall be taken before further patient imaging is
performed.
3.11.6 Documentation
Record the results in the Table MP-8 (Ghost Image Evaluation) and the Table MP-S (MEDICAL PHYSICIST'S
MAMMOGRAPHY QC TEST SUMMARY).