3.17 Artifact Evaluation
Chapter 3
3.17.5 Performance Criteria and Corrective Action
Performance Criteria: No signicant artifacts should be observed.
Corrective Action: If artifact is present on the image, rotate acrylic 180 degrees, if artifact moves, then it is on the
acrylic.
If artifact is still in the same place, remove the paddle and repeat exposure. If artifact is not present then it is on the
paddle.
If artifact is still present, conduct the same test with lm/screen.
If artifact is still present, conduct the test again with a di󰀨erent cassette.
If artifacts persist, the medical physicist should consult with the interpreting radiologist as to whether the artifact is
clinically tolerable.
The source of the problem should be identied and corrective action shall be taken on the component(s) that caused
the failure within 30 days of the test date. Patient examination may continue during this period.
3.17.6 Documentation
Record the results in the Table MP-14 (Artifact Evaluation) and the Table MP-S (MEDICAL PHYSICIST'S
MAMMOGRAPHY QC TEST SUMMARY).

Table of Contents

Other manuals for Konica Minolta CS-3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta CS-3 and is the answer not in the manual?

Konica Minolta CS-3 Specifications

General IconGeneral
BrandKonica Minolta
ModelCS-3
CategoryMedical Equipment
LanguageEnglish

Related product manuals