2.8 Visual Inspection
2.8.1 Test Objective
To ensure the Xpress Digital Mammography System is fully operational and ready to process clinical images.
2.8.2 Frequency
MONTHLY
2.8.3 Required Equipment
Non woven, lint free cloth
Anhydrous Alcohol
2.8.4 Procedure
z
Mammography X-ray unit Evaluation
1
Follow the guidance given in the "Mammography Quality Control Manual" (ACR, 1999) Radio-
logic Technologist section procedure 11.
z
X-ray Control Area Evaluation
1
Ensure the following items:
Patient viewing window is clean and unobstructed.
All "power on" indicators are functioning.
z
CR Console
1
Ensure the following items:
All lights, indicators are function property.
The monitor of the REGIUS Console CS-3/CS-7 is free of smudges and ngerprints.
z
Imaging cassette and plate
1
Open each REGIUS cassette and inspect the surface of the imaging plate.
The surface should be free of scratches, dust or other substances or damage.
2
Inspect the REGIUS cassette housing.
No cracks should be present.
Unlock mechanism should operate smoothly.
No dust or other particulate matter should be present.