3.13 AEC Performance Checks
3.13.1 Stability and Reproducibility
z
Test Objective
To evaluate the stability and reproducibility of the Automatic Exposure Control (AEC) system.
z
Frequency
Initially before patients are imaged
ANNUALLY
As part of an MEE
z
Required Equipment
RMI-156 phantom
Imaging cassette
Calibrated dose meter
z
Procedure
1
Register the patient on the REGIUS Console CS-3/CS-7.
2
Set the REGIUS Console CS-3/CS-7 for a minimum of four exposures under the AEC Checks
exam tag.
HINT
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
Use the same cassette for all exposures in this test.
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
3
Position the RMI phantom in the center of the breast support table at the edge of the chest
wall. The phantom should cover the cells used for AEC.
4
Place the detector of the dose meter on the breast support table next to the RMI phantom. It
should be 4 cm from the chest wall and not over the cells used for AEC.