3.5 Radiation Output Rate
4
Make three exposures and record the mGy and the exposure time in the appropriate elds on
Table MP-4 (Radiation Output Rate). The Table MP-4 will denote Pass/Fail.
3.5.5 Performance Criteria and Corrective Action
Performance Criteria: The system shall be able to produce a minimum output of 7.0mGy air kerma per second (800mR
per second) when operating at 28 kVp in the standard mammography mode (Mo/Mo).
Corrective Action: If the test fails, please repeat the above procedure. If results exceed the criteria the mammography
X-ray unit must be serviced to correct the condition.
The source of the problem shall be identied and corrective action shall be taken within 30 days of the test date. Pa-
tient examination may continue during this period.
3.5.6 Documentation
Record the results in the Table MP-4 (Radiation Output Rate) and the Table MP-S (MEDICAL PHYSICIST'S
MAMMOGRAPHY QC TEST SUMMARY).