3.5 Radiation Output Rate
Chapter 3
3.5 Radiation Output Rate
Table MP-4
Date: Inspectors Name:
* Measurements should be taken using a minimum 3 sec. exposure
oM/oM:retliF/tegraT
Measured tube voltage: 28kVp
Repeat #
Total
Radiation
Output mGy
Measured
exposure time
sec
Radiation
Output Rate
mGy/s
Rep #1
Rep #2
Rep #3 Pass/Fail
Mean
Criteria 7.0mGy air kerma/sec(800mR/sec) P Pass
F Fail
Comments:
Mammography X-ray unit S/N: