3.12 S value Response Indicator
Chapter 3
4
Wait 2 minutes and then process the cassette.
5
Record the mR and the displayed S value on table MP-9 (S value Response Indicator).
6
Repeat steps 1 through 5 three more times.
IMPORTANT
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
This test is sensitive to variations in beam quality (kVp, target and lter material). Use an AEC mode that varies only mAs,
and does not allow variations in kVp, target and lter material.
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
Maintain consistent positions of the detector and the ROI at all times.
Make sure that the ROI does not include the detector of the dose meter. If it does, move the detector away and do the test
over again.
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
3.12.5 Performance Criteria and Corrective Action
Performance Criteria: Each measured S value should be within +/- 20% of the value calculated using the formula of
S=2400/x [mR].
Where S is the displayed S value of the image and x is the value of the average exposure value stated in mR.
Corrective Action: If the test fails, repeat the procedure with a di󰀨erent cassette and plate pair. If the new pair passes
the test, patient examination may continue without using the cassette and plate pair that does not meet the perfor-
mance criteria.
If problems persist, contact your service representative and ensure that the uniformity and sensitivity calibrations are
completed. If problems still persist, the REGIUS system will need to have service to address the issue.
The source of the problem shall be identied and corrective action shall be taken within 30 days of the test date. Pa-
tient examination may continue during this period.
3.12.6 Documentation
Record this activity in the Table MP-9 (S value Response Indicator) and the Table MP-S (MEDICAL PHYSICIST'S
MAMMOGRAPHY QC TEST SUMMARY).

Table of Contents

Other manuals for Konica Minolta CS-3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta CS-3 and is the answer not in the manual?

Konica Minolta CS-3 Specifications

General IconGeneral
BrandKonica Minolta
ModelCS-3
CategoryMedical Equipment
LanguageEnglish

Related product manuals