3.12 S value Response Indicator
3.12 S value Response Indicator
Table MP-9
Date: Inspectors Name:
Cassette Bar Code #: CR Reader S/N:
Stability and Reproducibility
Exp Count
Measured
dose (mR)
S=2400/
X(mR)
(a)
Displayed
S value
(b)
S= 2400/ X(mR)
Exp #1
X(mR) is the value of the average exposure value stated in mR.
Exp #2
Exp #3
Exp #4
Mean Pass/Fail P Pass
% error of S (1-(a)/(b))
F Fail
Criteria S value variation; +/- 20%
Comments: