3.16 Spatial Resolution
3.16.5 Performance Criteria and Corrective Action
Performance Criteria: Evaluate the processed image looking for 7 LP/mm or greater.
Corrective Action: If this test yields results outside the action limits as specied, the source of the problem shall be
identied and corrective action shall be taken on the component(s) that caused the failure before further patient im-
aging is performed.
3.16.6 Documentation
• Record this activity in the Table MP-13 (Spatial Resolution) and the Table MP-S (MEDICAL PHYSICIST'S
MAMMOGRAPHY QC TEST SUMMARY).