126
3.16 Spatial Resolution
3.16.5 Performance Criteria and Corrective Action
Performance Criteria: Evaluate the processed image looking for 7 LP/mm or greater.
Corrective Action: If this test yields results outside the action limits as specied, the source of the problem shall be
identied and corrective action shall be taken on the component(s) that caused the failure before further patient im-
aging is performed.
3.16.6 Documentation
Record this activity in the Table MP-13 (Spatial Resolution) and the Table MP-S (MEDICAL PHYSICIST'S
MAMMOGRAPHY QC TEST SUMMARY).