4.2 MP-Tables
Table MP-1
Institution Name:
Date:
Inspectors Name:
Equipment
Room #:
Mammography X-ray unit S/N:
CR Reader S/N:
CR Console S/N:
Cassette Bar Code #:
Evaluation Item: Pass/Fail
snoitidnocgnitarepolamronrednuelbatssitinuehT-
- Assemblies move freely and cassette is secure in any position of the C-arm
- Support stands and compression paddles change normally
- Compression release and emergency stop function properly
- Compression thickness display is accurate within 5 mm of true value
- Operator and patient safety is not compromised during normal operation
- All lights, indicators, and audio signals function properly
P Pass
F Fail
Comments: