3.6 Average Glandular Dose
Chapter 3
5
The detector should be the same height as the top of the RMI phantom.
6
Expose the phantom and the detector of the dose meter with the kVp, target material and l-
tration at which the system is most commonly used clinically (Using AEC).
7
Make four (4) exposures and record the mR and mAs from the dose meter and console of the
mammography X-ray unit.
3.6.5 Performance Criteria and Corrective Action
Performance Criteria: Refer to "Mammography Quality Control Manual" (ACR, 1999) Medical Physicist section pro-
cedure 10.
Corrective Action: If the test fails, please repeat the above procedure. If results exceed the criteria the mammography
X-ray unit must be serviced to correct the condition.
The source of the problem shall be identied and corrective action shall be taken before further patient imaging is
performed.
3.6.6 Documentation
Record the results of the test in the Table MP-5 (Average Glandular Dose) and the Table MP-S (MEDICAL PHYSI-
CIST'S MAMMOGRAPHY QC TEST SUMMARY).

Table of Contents

Other manuals for Konica Minolta CS-3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta CS-3 and is the answer not in the manual?

Konica Minolta CS-3 Specifications

General IconGeneral
BrandKonica Minolta
ModelCS-3
CategoryMedical Equipment
LanguageEnglish

Related product manuals