EasyManuals Logo

Konica Minolta CS-3 Quality Control Manual

Konica Minolta CS-3
180 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #146 background imageLoading...
Page #146 background image
146
3.19 Phantom Image Quality (Medical Physicist)
IMPORTANT
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
The phantom image quality represents the total performance of the entire imaging chain. Changes in image quality may be
attributable to the mammography X-ray unit, REGIUS plates and cassettes, REGIUS Model 190/210/110HQ, laser printer,
lm, view box, or review workstation. Additional tests will be needed to identify the component(s), responsible for the change.
It is also necessary to ensure that the exam tag or the processing menu has not been changed. These factors will substan-
tially a󰀨ect the test results.
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
Mammography phantom images should always be viewed
By the same person
On the same viewing device
Under the same viewing conditions as the Radiologist review stations
Using the same type of magnication used for reading mammograms
Softcopy images should be window and leveled to best demonstrate the objects of interest
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
3.19.5 Performance Criteria and Corrective Action
Performance Criteria: Follow the guidance given in the "Mammography Quality Control Manual" (ACR, 1999) Medi-
cal physicist section procedure 7.
1
The S value result must not vary by greater than +/-20%.
2
The density di󰀨erence (disc vs. adjacent density) shall not vary by more than +/-0.05 from the
established operating level.
3
The optical density of the lm shall be within control limits of +/-0.20 from the established op-
erating level.
4
A minimum of (3) Specks, (3) Masses, and (4) Fibers should be visible.
5
The total number of simulated masses, speck groups, and bers visible in the phantom image
should not decrease by more than one-half, assuming the same individual is viewing the im-
ages under identical conditions.
6
The thickness displayed on the unit for the RMI Phantom should be accurate within +/- 0.5 cm
(3.7 cm-4.7 cm).
Corrective Action: If results are abnormal ensure proper thickness is displayed on the acquisition unit (4.2 cm for
RMI-156). Ensure REGIUS plate is clean. Repeat the test.
If problems persist, the source of the problem shall be identified and corrective action shall be taken on the
component(s) that caused the failure, before further patient examination is performed. If the problematic component
can be replaced by an alternative device, replace it and repeat the test. If the new combination passes the test, pa-
tient examination may continue using the system.
If the source of the problem is identied as a device used for image interpretation (i.e. laser printer, radiologist review
workstation) corrective action shall be taken before further mammographic image interpretation is performed by use
of that device. Patient imaging may be continued. If available, an alternative approved diagnostic device may be
used for mammographic image interpretation.

Table of Contents

Other manuals for Konica Minolta CS-3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta CS-3 and is the answer not in the manual?

Konica Minolta CS-3 Specifications

General IconGeneral
BrandKonica Minolta
ModelCS-3
CategoryMedical Equipment
LanguageEnglish

Related product manuals