3.15 CR Reader Scanner Performance
z
Jitter
1
Place a REGIUS cassette on the breast support table.
2
Place another REGIUS cassette into the cassette holder to activate the exposure.
3
Place the steel scale approximately in the center at right angles to the chest wall side of cas-
sette.
90°
4
Expose the cassette with 28 kVp 100 mAs (Mo/Mo).
5
Process the cassette on the breast support table, under the Geometric Distortion exam tag.