3.3 Tube Voltage Measurement and Reproducibility
3.3 Tube Voltage Measurement and Reproducibility
Table MP-2
Date: Inspectors Name:
Measured kVp value at the lowest measureable level
Rep #1
Displayed kVp value at the lowest measureable level
Rep #1
Pass/Fail
% error %( / -1)
Criteria +/-5%
Measured kVp at the most commonly used kvp setting Rep #1
Rep #2
Rep #3
Rep #4
Mean
Standard Deviation(SD) Pass/Fail
coeff of Variation (SD/Mean )
Criteria ≤2%
Displayed kVp at the most commonly used kVp setting Pass/Fail
% error %( / -1)
Criteria +/-5%
Measured kVp value at the highest available clinical kVp
Rep #1
Displayed kVp value at the highest available clinical kVp
Rep #1
Pass/Fail
% error %( / -1)
Criteria +/-5%
P Pass
F Fail
Comments:
Mammography X-ray unit S/N: