p/n 18603-001, Rev. C LTV
®
1200/1150 Ventilator Service Manual Page 8-3
Before removing the back panel
CAUTION
Opening the ventilator - Always turn the ventilator OFF and remove the external power
before opening the ventilator case or attempting to service the ventilator.
Anti-static precautions - Always wear a grounded anti-static wrist strap when handling the
ventilator with the case open. Electrostatic discharge (ESD) can damage ventilator
electronics.
Before replacing the back panel
Several checks must be made before reattaching the back panel. Follow the guidelines
listed under Back Panel, Reinstallation in this chapter.
After performing any maintenance
After performing maintenance on the LTV
®
1200/1150, the tests detailed in Chapter 9 - Final
Checkout Test must be run. Some maintenance procedures also require that the LTV
®
1200/1150 ventilator be partially or completely recalibrated.
CAUTION
Verification of operation - After opening the ventilator and performing any maintenance,
verify proper operation of the ventilator by performing the checks and calibrations
recommended in Chapter 9 - Final Checkout Test.