EasyManua.ls Logo

LTV 1150 - Appendix A - Ventilator Specifications; Appendix B - Glossary

Default Icon
341 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
p/n 18603-001, Rev. C LTV
®
1200/1150 Ventilator Service Manual Page B-1
Appendix B -
G
LOSSARY
TERM DEFINITION
AC
Alternating Current.
Airway Circuit
The airway tubing that connects the ventilator and the patient.
Airway Pressure
The airway pressure measured at the exhalation valve.
Airway Pressure
Display
A bar graph type display composed of 60 LEDs. This display shows the
real-time airway circuit pressure from –10 cmH
2
O to 108 cmH
2
O.
Alarm
An audible and visual notification that an alarm condition has been met.
Audible notification includes an oscillating or continuous tone. Visual
notification may include flashing displays, illuminated LEDs, and text
messages shown in the display window.
Apnea
Apnea occurs when the time between breath starts exceeds the set
apnea interval.
Apnea Backup
Ventilation
Apnea Backup Ventilation begins when an apnea alarm occurs and
continues until the patient initiates 2 consecutive breaths or the alarm is
canceled by an operator. Apnea Backup Ventilation is given in the
Assist / Control mode.
Apnea Interval
The maximum period of time allowed between breath starts. If the time
between breath starts exceeds this interval, an Apnea alarm occurs.
Assist / Control Mode
A mode of ventilation where the patient receives a minimum number of
machine and assist breaths. The available breath types are Volume
Control and Pressure Control.
Assist Breath
A volume or pressure breath that is initiated by the patient, and
controlled and cycled by the ventilator. Assist breaths may occur in
Assist / Control and SIMV modes.
Auto zero
The process of determining the transducer zero offset for ambient
pressure.
Bias Flow
A continuous flow of gas through the airway circuit during the
exhalation phase of the breath.
bpm
Breaths Per Minute.
Breath Period
The length of time between machine initiated breaths. The Breath
Period is determined by the Breath Rate setting. For example, a Breath
Rate of 6 would give a Breath Period of 10 seconds (60 seconds / 6
bpm).
Breath Rate, monitored
( f )
The number of breaths given in a minute, including machine, assist,
and patient breaths.
Breath Rate, set
The minimum number of machine breaths given in a minute.
BTPD
Body Temperature, Pressure Dry.
Circuit
See Airway Circuit.
Circuit Pressure
See Airway Pressure.
cmH
2
O
Centimeters of water pressure. A unit of measure for pressure.

Table of Contents

Related product manuals