Page 8-34 LTV
®
1200/1150 Ventilator Service Manual p/n 18603-001, Rev. C
Tubing Table
Tubing
P/N
Material
/Color
Diameter Length Qty Installed Location
0.62” (± 0.1”)
0.016m
2
One between elbow fitting from
high pressure port on valve diff.
transducer (PT7) & tee fitting and;
One between port #7 on solenoid
manifold & elbow fitting
1.10” (± 0.1”)
0.028m
1
Between port #1 on solenoid
manifold & tee fitting
1.75” (± 0.1”)
0.045m
2
One between port #9 (side port)
on solenoid manifold & tee fitting;
and
One between port #4 on solenoid
manifold & side panel low
pressure flow transducer port
2.06” (± 0.1”)
0.052m
1
One between port A on
accumulator & tee fitting
2.40” (± 0.1”)
0.061m
2
Between port #8 on solenoid
manifold & side panel high
pressure flow transducer port;
and
One between port #6 on solenoid
manifold & side panel exhalation
valve fitting
11834 Silicone
/Clear
0.188” (0.005m)
0.094” (0.002m)
O.D.
I.D.
3.25” (± 0.1”)
0.083m
1
Between port #3 on solenoid
manifold & airway pressure
transducer (PT6)
3.70” (± 0.1”)
0.094m
1
Between port B on accumulator &
PEEP pilot pressure transducer
(PT2)
10458 Silicone
/Clear
0.156” (0.004m)
0.094” (0.002m)
O.D.
I.D.
4.70” (± 0.1”)
0.118m
1
Between port #2 on solenoid
manifold & flow diff. transducer
(PT4)
0.50” (± 0.1”)
0.013m
3
One between tee fitting & bottom
leg of wye fitting; and
One between high pressure port
on valve diff. transducer (PT7) &
elbow fitting; and
One between low pressure port
on valve diff. transducer (PT7) &
elbow fitting
0.85” (± 0.1”)
0.022m
1
Between plug fitting & bottom
pilot-in port.
1.00” (± 0.1”)
0.025m
2
One between wye fitting & middle
pilot-in port; and
One between wye fitting & bottom
pilot-out port
1.60” (± 0.1”)
0.041m
1
Between tee fitting & top pilot-in
port
10456 Silicone
/Clear
0.125” (0.003m)
0.063” (0.002m)
O.D.
I.D.
1.75” (± 0.1”)
0.045m
1
Between middle pilot-out port &
tee fitting at side panel exhaust
vent slot