EasyManua.ls Logo

LTV 1150 - External Inlet Filter

Default Icon
341 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 8-62 LTV
®
1200/1150 Ventilator Service Manual p/n 18603-001, Rev. C
External Inlet Filter
The external inlet filter should be removed and cleaned once a month. If the ventilator is
being operated in high dust or humidity environments, it may need to be cleaned more often.
If the filter is damaged or can not be thoroughly cleaned, it should be replaced.
Parts required for replacement: Tools required:
Replace if damaged:
Inlet Filter, Reticulated Foam P/N 10258
Mild cleanser
Soft cleaning brush
To clean or replace the External Inlet filter:
1) Gently pinch the external inlet filter and remove it from its housing on the left side of the
ventilator.
2) Clean the filter using a mild cleanser and warm water. Rinse the filter thoroughly to
remove all traces of the cleanser and gently wring it out. Allow the filter to dry completely
before replacing it in the ventilator.
3) Inspect the filter for damage. If the filter is not intact, shows signs of damage or cannot
be completely cleaned, replace it with a new filter.
4) Replace the filter by tucking it into its housing on the left side of the vent. Make sure the
filter lies flat and is seated all the way into the housing.

Table of Contents

Related product manuals