Page 9-30 LTV
®
1200/1150 Ventilator Service Manual p/n 18603-001, Rev. C
12 Hour Burn-in Worksheet
LTV
®
1200/1150 SERIAL NUMBER:
____________
CONDUCTED BY:
____________
MEMORY BOARD SOFTWARE VER.:
____________
DATE:
____________
TEST
DESCRIPTION
STEP
#
MEAS.
VALUE
REQUIREMENT
PASS /
FAIL
12 Hour Burn-in
Connect AC adapter and test lung. 1)
Power ventilator up. 2)
Set the ventilator controls to the settings
listed.
2)
Run the ventilator connected to a test
lung for 12 hours.
3)
Verify that no alarms occur.
Return the ventilator controls to the
factory default settings.
4)