p/n 18603-001, Rev. C LTV
®
1200/1150 Ventilator Service Manual Page A-1
Appendix A -
V
ENTILATOR
S
PECIFICATIONS
Modes and Breath Types
Breath Types Volume Control, Pressure Control, Pressure Support, Spontaneous
Modes Control, Assist/Control, SIMV, CPAP, NPPV, Apnea Backup
Variable Controls
Control Range Tolerance
Backup Pressure Trigger -3 cmH
2
O
± 2 cmH
2
O
Breath Rate “--”, 1 to 80 bpm
± 1 bpm or 10% of breath period,
whichever is less
Date Format mm/dd/yyyy, dd/mm/yyyy,
yyyy/mm/dd
n/a
Display Select Toggles between manual or
automatic display scrolling
and changes monitor
displayed.
n/a
One push toggles monitor
window display between
normal display, INSP HOLD
and EXP HOLD.
While INSP HOLD is
displayed, a push and hold
initiates an Inspiratory Hold.
6 seconds maximum
Inspiratory/Expiratory
Hold
While EXP HOLD is
displayed, a push and hold
initiates an Expiratory Hold.
6 seconds maximum
Inspiratory Time 0.3 to 9.9 seconds
± 0.05 seconds
Leak Compensation On, Off n/a
Language English, Dansk, Deutsch,
Espanol, Francais, Italiano,
Norsk, Portugues, Svenska,
Русско
n/a
O
2
% (LTV
®
1200 Only) 21% to 100%
O
2
% mean: 21% to 50%: ± 3
51% to 100%: ± 5
keep steady-state only
O
2
: 95%
± 5
(O
2
Flush) (LTV
®
1200
Only)
Time: 1, 2, or 3 minutes
± 0.1 sec