p/n 18603-001, Rev. C LTV
®
1200/1150 Ventilator Service Manual Page 2-3
Alarm Test
The alarm test verifies that the audible alarm is working correctly.
To run the alarm test:
1) Press Select while ALARM is displayed.
2) Verify that the audible alarm sounds. When the alarm has sounded for at least 2
seconds, press Select again.
• The audible alarm is silenced and the next menu item is displayed.
3) Verify a confirming audible chirp sounds after the alarm is silenced.
If the ventilator fails the alarm test, see Chapter 7 - Troubleshooting for more information.
NOTE
Although the LTV
®
1200 Front Panel is shown, the test is applicable to the LTV
®
1150
Ventilator.