Page 6-30 LTV
®
1200/1150 Ventilator Service Manual p/n 18603-001, Rev. C
Solenoid
The Solenoid menu is used to test the operation of the solenoids. The menu is set up as
follows:
SOLENOID
ExhPilot OFF
Purge OFF
APres OFF
ExhDiffP OFF
PPin OFF
PPout OFF
O2 #1 OFF (LTV
®
1200 only)
O2 #2 OFF (LTV
®
1200 only)
O2 #3 OFF (LTV
®
1200 only)
O2 #4 OFF (LTV
®
1200 only)
EXIT
Use the Set Value knob to rotate between solenoid options. To change the state of any
solenoid, press Select while it is displayed in the window. Pressing Select toggles the state
of the solenoid on and off.
Solenoids are automatically set to the off state when you enter the Calibration menu.