p/n 18603-001, Rev. C LTV
®
1200/1150 Ventilator Service Manual Page 9-29
Extended Features, Default Settings
Feature Default Feature Default
Alarm Volume 85 dBA Patient Query On
Apnea Interval 20 sec Patient Size INFANT
Com Setting Data PC Flow Term Off
Control Unlock Easy PIP LED On
Date Format MM/DD/YYYY Rise Time Profile 4
High f Alarm High f Off SBT Display f/Vt On
High f Alarm Delay 30 sec SBT FIO
2
80
21%
High PEEP Alarm PEEP + 5 cmH
2
O SBT High f Alarm 35 bpm
HP Alarm Delay No Delay SBT High f/Vt Alarm 105 bpm / l
Language English SBT Low f Alarm 10 bpm
Leak Compensation On SBT Low f/Vt Alarm 70 bpm / l
Low PEEP Alarm PEEP - 5 cmH
2
O SBT Mode Off
LPP Alarm All Breaths SBT Mode Run Time 20 min
NPPV Mode Off SBT PEEP 0 cmH
2
O
O
2
Duration Cylinder
79
2000 psi or 138 bar SBT Pressure Support 10 cmH
2
O
O
2
Duration Cylinder
79
622liters Var. Flow Term 25%
O
2
Flush Period
79
3 min Var. Time Term 1.5 sec
Patient Assist Normal
3) Run the ventilator connected to a test lung for 12 hours and verify that no alarms occur.
4) Return the ventilator settings to factory defaults (see page E-11) at the conclusion of the
test.
79
Available on the LTV
®
1200 Only.