Page D-8 LTV
®
1200/1150 Ventilator Service Manual p/n 18603-001, Rev. C
Event Name Code Event Associated Alarm
STACK 56 Stack overflow detected RESET
SYNCER0 37 Stepper motor lost sync exited HW FAULT
SYNCER1 36 Stepper motor lost sync occurred HW FAULT
TBN ESTP 46 Turbine emergency stop occurred HIGH PRES
TBN HSTP 66 Turbine Hold Stop occurred HIGH PRES
TBN ISTP 44 Turbine immediate stop occurred HIGH PRES
TBN ZERO 45 Turbine set to zero flow occurred HIGH PRES
VENT 0 02 Power off None
VENT 1 01 Power on None
VENT CHK 04 Set vent check Entered VENT CHECK mode
WDOG TST 59 Watchdog test run Vent Inop
XDC FLT0 20 XDCR fault exited XDCR FAULT
XDC FLT1 19 XDCR fault occurred XDCR FAULT
XDCR AIR 64 Airway pressure transducer fault XDC FLT1
XDCR BI 63
Differential pressure transducer -
Bi-directional channel fault
XDC FLT1
XDCR NAR 61
Differential pressure transducer -
Narrow channel fault
XDC FLT1
XDCR WID 62
Differential pressure transducer -
Wide channel fault
XDC FLT1
N/A
55
Not used
N/A
30
Not used
N/A
29
Not used
N/A
07
Not used