p/n 18603-001, Rev. C LTV
®
1200/1150 Ventilator Service Manual Page 4-5
4) Inspect the filter for damage. If the filter is not intact, shows signs of damage or cannot
be completely cleaned, replace it with a new O
2
Inlet Filter (P/N 14313) and O-Ring (P/N
10609), available from Pulmonetic Systems.
5) Replace the filter by sliding it back into the O
2
inlet port. Replace the O-Ring, making
sure it is completely tucked under the retaining lip on the inside of the O
2
inlet port.
6) Reconnect the high pressure O
2
line.
CAUTION
Wet or Damp Filters - Do not install a wet or damp filter into the LTV
®
1200 ventilator. This
could damage the ventilator.