p/n 18603-001, Rev. C LTV
®
1200/1150 Ventilator Service Manual Page E-15
General Tools:
The following general tools are required to perform various procedures on the LTV
®
1200/1150 ventilator;
• Air supply, (0-50 cmH
2
O)
• Allen wrench, 7/64”
• Amp meter, 10 amp to 60 Hz
• Compressed O
2
source (0-80 PSI regulator)
(LTV
®
1200)
• Compressed gas source (0-50 cmH
2
O
regulator)
• Dental pick
• Drill Bit, 3/16”
• Drill, electrical
• Multi-Meter, Digital
• Nut drivers, 3/16”, 1/4”, 1/2” & 13mm adapters
for torque wrench
• O
2
Analyzer (calibrated)
• O
2
Supply (0-80 PSIG) (LTV
®
1200)
• O
2
Inlet connector (LTV
®
1200)
• Pliers, needle nose
• Pop Rivet tool
84
• Pressure gauge (FSD >80 PSI) (LTV
®
1200)
• Pressure manometer (0-100 cmH
2
O)
• Screwdriver, Phillips with torque meter
• Screwdriver, Straight tip
• Spirometer (calibrated)
• Test Lung (2 Liter or other large-compliance
reservoir) or
• Test Lungs (2 x 1 liter or greater)
• Tie wrap tool
• Torque Screwdriver, 120 in-oz.
• Variable DC voltage source (10 Amp)
• Wrist strap, grounded anti-static
• Y-connector for test lungs
Unique Tools:
The following unique tools are required to perform various procedures on the LTV
®
1200/1150 ventilator and are available from Pulmonetic Systems separately, or as part of
the Maintenance and Calibration Kit, P/N 11566;
84
Pop Rivet tool capable of setting .114” diameter shaft, Pop Rivets.
85
DISS/ NIST adapter required to comply with some international gas fittings.
86
In the European Union, Loctite® 8104 may be substituted as an equivalent compound.
87
O
2
Sampling Tube, P/N 10544, ~10.0” long, 0.125” O.D. X 0.079 I.D. clear polycarbonate tubing.
• Calibration Syringe Assembly
P/N 11471
• DISS to NIST fitting for high pressure O
2
port
P/N 10702
85
• Driver, Torque Dial Indicator
P/N 11574
• External Battery Test Cable Assembly
P/N 11474
• Flow Valve Insertion Tool
P/N 14206
• Internal Battery Test Cable Assembly
P/N 11472
• Lubricant, Silicone Compound
P/N 10123
86
• O
2
Sampling Tube
P/N 10544
87
• Patient Assist Cable, Normally Closed
P/N 10779
• Patient Assist Cable, Normally Open
P/N 10780
• Pinch Clamp
P/N 11529
• Power PCB Removal Tool
P/N 11599
• Service Cable Assembly
P/N 11485
• Stepper Motor Calibrator
P/N 10871
• Tube Adapter, 22mm
P/N 10696
• Turbine Pressure Test Adapter Assembly
P/N 11567