Series F Test Frames + IntelliMESUR
®
Software User’s Guide
64
TOC
Zero the indicator associated with the master load cell. Engage the calibration fixtures, and
ensure that there is no load applied to the force sensor. Then select Next.
9.
Apply the indicated compression load using the crosshead motion controls. The load value
defaults to even intervals. For example, for a 50 lbF force sensor with 5 calibration points, the first
default load value is 10 lbF. All load values may be changed by up to 20% in either direction.
Select Next. Follow the same procedure for all required loads.
10. After the final compression load, the screen appears as follows:
Configure the calibration fixtures and equipment for tension loading. Then select Next.