Operating Instructions – Labmaster
©
130 and MB20G Systems
Instruction Manual
Chapter 9: System Operation
© MBraun, Inc. Edition 1/2007 Page 9-9
Touch the Evacuate/Start Autom. button to
evacuate the antechamber of room atmosphere.
Wait for the gauge to settle at the negative
position and continue to evacuate for:
Non-porous Material: 5 minutes
Porous Material: 12 hours
Touch the Evacuate/Start Autom. button to
deactivate the evacuation process.
Touch the Refill button to refill the antechamber
with box gas.