2.3.1 For pumps with 9V alkaline battery
Applies to models: Rythmic™Evolution, Mini Rythmic™ Evolution, Rythmic™
Evolution Organiser 500/100.
To access the battery please follow the below procedure:
- Verify that the pump is switched OFF.
- Place the pump on a table, with the front panel facing downwards.
- Open the transparent covers as described in chapter 2, paragraph 2.1
"Opening the transparent covers".
2.3 Power Supply
16 17
!Warning
•
Do not use rechargeable NiCd or NiMH batteries. Do not use carbon zinc batteries.
•
Ensure that the battery placed in the pump is in good condition and the remaining
battery power is enough for the infusion that is going to start. If the battery is not likely to
finish the infusion, replace it before infusion starts. Loss of power may result in non-delivery
of medication which could result in death or serious injury to patient.
• Always have a new battery available for replacement to ensure that the pump can
continue its operation for the time specified. Loss of power may result in non-delivery of
medication which could result in death or serious injury to patient.
• Ensure the battery door has been locked in position. If the battery door becomes loose,
then the battery can be removed, which could result in loss of power and non-delivery of
medication which could result in death or serious injury to patient.
• Always inspect the battery compartment for fluid or debris before inserting the batteries.
Fluid or debris in the battery compartment may damage the battery terminals and could
result in loss of power and non delivery of medication.
• Always inspect the battery terminals for debris, grime and damages. If debris or grime
exists, clean the terminals according to cleaning instructions. If the debris or grime persists
or damage exists remove the device from service. Debris, grime or damage of the battery
contacts could result in loss of power and non delivery of medication.
Insert the batteries within
the compartment, observing
the correct polarity as shown
by the illustration within
the battery compartment.
Slide the battery
compartment lid in the
direction of the arrow
marked BATT.
Slide the lid back
in place.
IEC 6LR61
IEC 6LR61