2.6
Fitting the Administration set into the pump
Sets with reservoir, Rythmic™ Full Sets
Rythmic™ Full Sets can only fit in Rythmic™ and Mini Rythmic™ Evolution models.
1. Open the transparent cover.
2. Press the tubing cover release lever to open the cover. This will expose the
pumping mechanism.
3. Remove empty administration set, if applicable.
4. Load the new bag into the pump with the filling port of the bag oriented nearby
the key lock of the transparent cover.
!Warning
•
Ensure that the slide clamp is closed before removing the administration set from the
pump, to prevent free flow of medication.
•
Always inspect for any marks, dust, dirt, particles or substance underneath the tubing
cover or in the mechanical pumping area before use. Any of the above could result to
under or non-delivery of medication. The pump should be immediately taken out of use
and inspected by qualified service personnel to ensure its proper function prior to reuse.